RD27 note 20, 14 December 1993

FORWARD CALORIMETRY AND THE LEVEL-1 TRIGGER
 
N. Ellis
PPE Division, CERN

Introduction

The designs for the level-1 calorimeter trigger that have been performed so far in RD27 have concentrated on the electron/photon and jet triggers, although the need for a missing transverse energy trigger has always been recognised. The forward calorimeters, typically covering the pseudorapidity range |eta| = 3.0 - 4.5, are essential for the missing energy measurement {1}. We must therefore make provision for sending data from these devices to the level-1 calorimeter trigger processor.

It is worth noting that in ATLAS it is not planned to recompute the missing transverse energy vector at level-2. However, high-ET objects that contribute to the missing transverse energy can be validated and some "corrections" could be made to the missing transverse energy vector.

Various technologies are under consideration for the forward calorimeters. In ATLAS, a choice will be made between high-pressure gas, liquid argon and liquid scintillator calorimeters. The number of cells in these calorimeters is typically ~2000 with two depth samplings. The noise per calorimeter cell may be as high as 1-2 GeV energy equivalent, but it should be kept in mind that sin q < 0.1 so that the noise expressed in terms of transverse energy should be manageable.

The forward calorimeters may be used to identify forward-going jets in the physics analysis ("jet tagging"). However, it does not seem necessary to include this information in the trigger.

Connection of Forward Calorimeters to Level-1 Trigger

As discussed above, the only use of the forward calorimeters in the level-1 trigger is for the missing transverse energy trigger. There is therefore no need for the complicated EM cluster logic (and jet logic) that is used in the central calorimetry. However, the most economical solution may be to interface the forward calorimetry to the standard trigger processor inputs, even if much of the trigger logic is not used. In the bit-parallel design [1], the processor receives zero-suppressed data, tagged by the bunch-crossing number, over asynchronous serial links; in the bit-serial design [2], high-speed synchronous serial links are used.

The data received by the level-1 calorimeter trigger processor is transverse energy (not total energy). After partial summing, the ET data are converted using look-up tables to Ex and Ey (components of the transverse energy vector). Extending the arguments of [2], one should preserve reasonable granularity in azimuthal angle, phi, at the input to the processor. However, one can sum ET over eta externally without loss of information for the missing transverse energy calculation.

Two issues that must be considered when summing ET before the input to the trigger processor are the following:

Two possibilities are being considered for the calorimeter readout:

Interface to the level-1 trigger with FERMI readout for the calorimeter

If FERMI is used for the calorimeter readout, with cell thresholds applied at the finest granularity, one could perform digital summation over the full eta region, giving one input per phi region (per end plug) to the trigger processor. For Df = 0.1, this would require 64 processor inputs, corresponding to one "cluster-finding" module (per end plug) in the bit-parallel processor design.

The first part of the summation would be done on the FERMI microsystems, with further summation on the FERMI boards. One could imagine completing the digital summing tree on forward- calorimeter trigger interface modules, which would also contain the standard interface to the trigger processor. In the case of the bit-parallel design, this would include zero-suppression, tagging, parallel-to-serial conversion and electrical-optical translation. This interface would be connected to the trigger processor via an optical link.

Interface to the level-1 trigger with analogue-pipeline readout for the calorimeter

If a readout system with analogue pipelines is adopted for the calorimeter readout, the trigger signals will have to be formed by analogue summation. Digitisation will then be performed by dedicated trigger ADCs followed by look-up tables to apply calibration constants giving transverse energy output and perform trigger cell thresholding. Here the following considerations apply: Digitisation could be performed using standard trigger ADC modules, incorporating the interface to the trigger processor, in which case a relatively large number of processor inputs may be required. Alternatively, one could follow the ADC / look-up table system with a digital summing tree as in the FERMI based system discussed above.

A limitation of the analogue summing scheme described here is that it may not be possible to adjust the weights in the analogue summation (or to monitor the calibration of individual calorimeter channels contributing to the trigger). It would therefore be important to have a stable calorimeter response.

Conclusions

Forward calorimetry should be included in the level-1 calorimeter trigger for use in the missing transverse energy trigger. This should be possible whether FERMI or an analogue pipeline solution is adopted for the calorimeter readout, although there are some advantages in using FERMI. In either case, it will be necessary to build forward- calorimeter interfaces to the level-1 calorimeter trigger processor - these could make use of standard components (e.g. trigger ADC, interface to optical data link).

References

[1] RD27 note 8.
[2] RD27 note 6.
[3] FERMI (RD16) status report: CERN/DRDC/93/21.
----
Footnote (1):
If the forward-calorimeter information were ignored, the missing
transverse energy trigger would be dominated by
events with high-pT jets falling outside  he
the acceptance of the central calorimeters.